1.General Description This Colony Counter is a semi-automatic digital display for counting the germs, which consists of counter, sensor, counting cell etc. The counter is made of CMOS integrated electric. In the black longitudinal background counting cell, a high performance flex LED strip is...

Send InquiryChat Now
Product Details

1.General Description

This Colony Counter is a semi-automatic digital display for counting the germs, which consists of counter, sensor, counting cell etc. The counter is made of CMOS integrated electric. In the black longitudinal background counting cell, a high performance flex LED strip is used for side lighting, which can give a clear colony comparing. The monitor is designed as three-figure number according to the requirement for bacterial counting SOP. When the colony in one culture plate is over 300, dilute the sample and redo the operation in order to ensure the accuracy of the counting. This instrument can reduce the labor of the operator , enhance the efficiency and working quality, which can be widely used in the field of foodstuff, beverage, drugs, biological products, sanitary utilities, drinking water, industrial waste water and clinical sample. It is a necessary instrument for all levels of health and epidemic prevention stations, environmental monitoring stations, food sanitary supervision organizations, hospitals, biological products manufacturers, medicine inspection stations, foodstuff companies, chemical industry, institutions, and research centers.

2.Main Data

2.1Capacity of the Counter: 0~999

2.2Power consumption of lamp: 9w

2.3Total power consumption: <12w

2.4Voltage : AC100-240V, 50/60Hz

2.5Volume: 255×210×160 mm

2.6Weight: 2.2kg

3.Instructions

3.1 Connect the power, open the lamp in the counting cell when the monitor shows“001”, insert the sensor and push the “reset” key.

3.2 Put the sample into culture plate.

3.3 Press the red button on the machine side for opening the light under counter plate. You can also adjust suitable brightness by screwing the side knob.

3.4 If the brightness of the under light is not enough for observation, just press the button beside the magnifier to open the upper light(Double LED lamp with four 11.6mm button cells supplying power, standard configuration). Do pull out the insulating plastic strip when using the upper light at first time.

3.5 Count the colony in the plate with the sensor one by one. At this time, the colony is colored and buzzer sounds which means the counting is ok. On the monitor, the numbers are added automatically.

3.6 Check with magnifier to ensure the counting is correct.( If  error  happening  when counting by the touching pen, continue the counting by pressing △ or ▽)     

3.7 The last number in the monitor is the colony in the plate.

3.8 After the operation, please take out the culture plate. Push the   “reset” key and the monitor return to original state.

4.Notice

4.1 The instrument should be put on a fixed and stable operation table with smooth face.

4.2 When counting the colony, the sensor should not be over leaned. With a slight touch to a sense of elastic, the number is input.

4.3 Far from the humidity, severe shock, strong sunshine, acids and alkali. A dust cover is necessary.

4.4 Prevent the culture media from contaminating the counting cell.  

4.5 The instrument and the sensor should not be dismantled at random. If the trouble occurs, please ask the experienced technician for repairing.    


We are professional as one of the leading j-2 colony counter manufacturers and suppliers in China. Welcome to wholesale the high quality and durable laboratory equipment from our factory. And good after-sale service and timely delivery are offered.