How To Use High Pressure Steam Sterilizer And Matters Needing Attention

Leave a message

1. How to use the high pressure steam sterilizer


(1) Add an appropriate amount of water to the outer pot, put the items to be sterilized into the inner pot, cover the pot and tighten the screw symmetrically.


(2) Heating will generate steam in the pot. When the pointer of the pressure gauge reaches 33.78kPa, open the exhaust valve to discharge the cold air. At this time, the pointer of the pressure gauge will drop. When the pointer drops to zero, the exhaust valve will be closed.


(3) Continue heating, the steam in the pot increases, and the pointer of the pressure gauge rises again. When the pressure in the pot increases to the required pressure, reduce the fire power, and keep the steam pressure at the required pressure for a certain period of time according to the characteristics of the sterilized items. , and then turn off the power or fire of the sterilizer, let it cool down naturally, and then slowly open the exhaust valve to remove the residual air, and then open the cover and take out the food.


2. Precautions for high pressure steam sterilizer:


(1) The items to be sterilized should not be placed too tightly.


(2) The cold air must be fully removed, otherwise the temperature in the pot will not reach the specified temperature, which will affect the sterilization effect.


(3) After sterilization, do not deflate and decompress, otherwise the liquid in the bottle will boil violently, flush out the cork and overflow or even cause the container to burst. The lid can only be opened after the pressure in the sterilizer is reduced to equal to the atmospheric pressure. Also known as: autoclave, autoclave, autoclave, portable autoclave, portable autoclave, portable autoclave, portable autoclave, portable autoclave Sterilizer, Portable Autoclave


(4) The test tube or bottle containing the medium should be wrapped with oiled paper or kraft paper to prevent condensed water from entering.


(5) In order to ensure the sterilization effect, the sterilization effect should be checked regularly. The commonly used method is to put sulfur powder (melting point of 115°C) or benzoic acid (melting point of 120°C) in a test tube, and then conduct a sterilization test. If the above substances are melted, it means that the temperature in the high pressure steam sterilizer has reached the requirement, and the sterilization effect is reliable. It is also possible to stick the adhesive tape (with a temperature-sensitive indicator on it) to detect the effect of the sterilizer on the outer packaging of the item to be sterilized. If the indicator on the adhesive tape changes color, it also means that the sterilization effect is reliable.


3. Application of high pressure steam sterilizer:


It is a commonly used sterilization method, and it is generally treated at 101.33kPa for 15 to 20 minutes, which can achieve the purpose of sterilizing articles. All items that are resistant to high temperature and humidity, such as common culture medium, physiological saline, clothes, gauze, glass equipment, etc., can be sterilized by this method.